NEW! Eltra – Oxygen/Nitrogen Analyzer – Elementrac ON‑p 2
The ELEMENTRAC ONH-p determines oxygen, nitrogen and hydrogen in inorganic samples by inert gas fusion in an impulse furnace with temperatures in excess of 3.000 °C.
Technical Specifications
Measured elements | nitrogen, oxygen |
Samples | inorganic |
Furnace alignment | vertical |
Sample carrier | graphite crucibles |
Field of application | ceramics, engineering / electronics, steel / metallurgy |
Furnace | electrode impulse furnace (max. 8,5 KW*), temperatures in excess of 3,000 °C |
Detection method | solid state infrared absorption for oxygen thermal conductivity for nitrogen |
Typical analysis time | 120 – 180 s |
Chemicals required | copper oxide, magnesium perchlorate, sodium hydroxide |
Gas required | compressed air, helium 99.995 % pure, argon 99.995% pure (if required), all gases with (2 – 4 bar / 30 – 60 psi) |
Power requirements | 3~ 400 V, 50/60 Hz, max. 8,500 W |
Dimensions (W x H x D) | 57 x 77 x 63 cm |
Weight | ~ 161 kg |
Required equipment | PC, monitor, balance (resolution 0.0001g) |
Optional accessories | carrier gas purification, external chiller, gas calibration unit |
* limited to 6.8 kw in application settings |
Key Features
Elementrac ONH-p
- Low gas consumption and high sensitivity due to closed gas system
- Easy application of pins, powders and granules
- Inexpensive argon as carrier gas possible
- Short analysis time
- Powerful impulse furnace with 8.5 kW
- Optional Autocleaner
- Reliable ON elemental analysis of inorganic samples like steel, non-ferrous metals, ceramics, slags, ores, etc.
Operation and Analysis Process
Step 1: Logging the sample into the ELEMENTS software
The sample ID is logged into the software and the weight is automatically transferred (see step 2).
Step 2: Weighing and introduction of sample into the port
The ELEMENTRAC OH-p 2 analyzes volumes from a few mg up to 2 grams safely and precisely. Rod-shaped or granular samples can be applied directly. For the elemental analysis of powders, a capsule is recommended which does not have to be sealed.
Step 3: Analysis
The empty graphite crucible is then placed on the lower electrode and the elemental analysis is started via the ELEMENTS software. The software controls all subsequent process steps.
Step 4: Data output and export
120 to 180 seconds after the analysis has started, the measured concentrations are available for export as a report or via LIMS.
Oxygen, Nitrogen and Hydrogen Analyser ONH-p 2 Introduction
Key Industries
- Construction
- Machinery & Manufacturing
- Mining & Minerals
Brand
ELTRA is a specialist for elemental analyzers for the accurate and rapid analysis of solid materials. ELTRA equipment is a standard for elemental analysis in industries such as mining, steel, energy, cement and metal working.
ELTRA analyzers are used to reliably determine elements like carbon, hydrogen, sulfur, oxygen and nitrogen, and cover a wide, dynamic measuring range. The ability to adapt each analyzer to user specifications allows the precise measurement of a great variety of organic and inorganic samples in a multitude of applications.