Eltra – Carbon/Hydrogen/Sulfur Analyzer – CHS-580

  • LAOS

    DKSH Laos Company Ltd.
    3rd Floor, Vientiane Center Office Building Khouvieng Road,
    Sisattanak District, Vientiane Capital, Laos

    +856 21 453 100

    +856 21 453 555

    Email

  • CS-580, CHS-580 Brochure


ELTRA‘s CHS-580A is the ideal elemental analyzer for simultaneous determination of carbon, hydrogen and sulfur in organic samples. It features a resistance furnace with vertical ceramic tube and uses ceramic crucibles as sample carriers.

An automatic sample loader allows for analysis of large numbers of samples without operator intervention. The sample loader is available in two versions to accommodate 36 or 130 crucibles. For reliable analysis of acidified samples (TOC analysis), both versions are available in an acid-resistance configuration.

The temperature can be set in steps of 1 °C to a maximum of 1,550 °C. Thanks to sample weights of up to 500 mg and more even inhomogeneous materials can be reliably analyzed. The analyzer can be equipped with up to four independent infrared cells for measuring carbon and sulfur concentrations, allowing for a great variety of applications.

Application Examples

Coal, coke, oil, plant materials, rubber, soot, tobacco, waste
 

Function Principle

Operation CHS-580

After weighing the sample in a ceramic boat, the weight is transferred from the interfaced balance to the PC. If required, sample weights can also be entered manually. Then the ceramic boat is introduced into the furnace for combustion. The average analysis time is 60 to 180 seconds. The detector signals and instrument parameters are displayed during analysis. Evaluation of the signals and display of the results are done automatically; the data can be transferred to a laboratory information management system (LIMS). The CS-580 requires minimum maintenance. The particle filters and chemicals which need to be maintained are easily accessible.
 

Measuring Principle CHS-580

In the CHS-580 the sample is burnt in an oxygen atmosphere at temperatures up to 1,550°C. The furnace temperature can be freely selected in steps of 1°C. The combustion gasses (CO2, H2O, SO2) coming from the furnace and first pass through a dust filter and then into the heated H2O infrared cell. After the water vapor is chemically absorbed, the dried CO2 and SO2 gas is detected in the additional infrared cells. Depending on the configuration, it is possible to combine up to three infrared cells with different sensitivities.

Technical Specifications

Measured elementscarbon, hydrogen, sulfur
Samplesorganic
Furnace alignmenthorizontal
Sample carrierceramic boats
Field of applicationagriculture, biology, chemistry / plastics, coal / power plant, construction materials, environment / recycling, medicine / pharmaceuticals
Furnaceresistance furnace (ceramic tube), adjustable up to 1550 °C (steps of 1 °C)
Detection methodsolid state infrared absorption
Number of IR cells1 – 3
Material of IR pathgold
Typical analysis time60 – 180 s
Chemicals requiredmagnesium perchlorate, sodium hydroxide
Gas requiredoxygen 99.5 % pure (2 – 4 bar / 30 – 60 psi)
Power requirements230 V, 50/60 Hz
max. heat up current 2000 Wmax. heat up current 2000 W
Dimensions (W x H x D)55 x 80 x 60 cm
Weight~ 70 kg
Required equipmentbalance (resolution 0.0001g), monitor, PC
Optional accessoriesTIC module, voltage stabilizer 5 KVA

Key Features

CHS-580

  • Simultaneous carbon, sulfur and hydrogen determination with minimal sample preparation
  • Wide range of organic materials can be analyzed
  • Rapid, precise, accurate and reliable element determination
  • Resistance furnace can be set up to 1550 °C in steps of 1 °C
  • Customized infrared cells provide wide, dynamic measuring range
  • Due to gold IR path, increased cell live time for analysis of halogen or acid containig samples
  • Powerful software (multilingual, customized display, export of results)
  • Single and multipoint calibration
  • Low maintenance
  • Robust design allows usage in production control and laboratory

Product Advantages

  • Fast simultaneous determination of CHS
  • Powerful horizontal resistance furnace up to 1,550 °C, adjustable in steps of 1 °C
  • High sample weights of 500 mg and more possible
  • Customizable measuring range
  • Individual infrared cell configurations available

Key Industries

  • Construction
  • Energy
  • Environmental
  • Mining & Minerals

Brand

Eltra

ELTRA is a specialist for elemental analyzers for the accurate and rapid analysis of solid materials. ELTRA equipment is a standard for elemental analysis in industries such as mining, steel, energy, cement and metal working.

ELTRA analyzers are used to reliably determine elements like carbon, hydrogen, sulfur, oxygen and nitrogen, and cover a wide, dynamic measuring range. The ability to adapt each analyzer to user specifications allows the precise measurement of a great variety of organic and inorganic samples in a multitude of applications.